Electrical Requirements
- Supply Voltage:
- 50 Hz: 200 – 240 V
- 60 Hz: 208 – 230 V
- Operating Current:
- 50 Hz: 17.0 A (@240 V)
- 60 Hz: 15.7 A (@230 V)
- Operating Power:
- 50 Hz: 3.4 kW (@240 V)
- 60 Hz: 3.6 kW (230 V)
- Supply Fuse Rating: 20 A (Starting Current: 65 A)
Specifications
- Heat Dissipation: Fan on enclosure outputs ~300 W of heat into the environment
- Service Requirements: Adsorber change @ approx. 30,000 hours
- Weight and Dimensions: H: 639 x D 610 x W 540 MM / 100 KG
- Water Cooling Requirements: 3.0 kW Chiller with typical flow rate of 5 l/m at 180C